The PRIMEDIC HeartSave myPAD 670A is a compact, lightweight fully automatic AED with voice prompts and illuminated guidance symbols. It analyses the heart rhythm and automatically delivers a shock when required, without the rescuer pressing a shock button. It includes adult/child modes using the same pads and IP66 dust and water protection, making it a convenient portable option. Designed for portability and ease of use, it is great for vehicles, toolboxes or kitbags.
GST-free.
| ARTG ID | 512488 — HeartSave myPAD 670A, non-rechargeable public automated external defibrillator |
|---|---|
| Operation | Fully automatic AED; analyses the heart rhythm and automatically delivers a shock when required, without a shock-button press |
| Defibrillation waveform | Biphasic truncated exponential (BTE), automatically compensated for patient impedance |
| Adult energy | 150 J → 170 J → 200 J, automatically escalating |
| Child energy / selection | 50 J; child-mode button |
| Analysis to shock readiness | With Battery 3C: 8 s to 150 J; 12 s to 200 J. With Battery 3G: 5 s to 150 J; 8 s to 200 J. |
| Device start to shock | With Battery 3C: 17 s to 150 J; 22 s to 200 J. With Battery 3G: 13 s to 150 J; 16 s to 200 J. |
| Asystole threshold | <100 µV |
| ECG sensing | Lead II; patient impedance 25–200 Ω; heart rate 30–300/min |
| Visual guidance — model 670A | Graphical instruction icons with LEDs; no display screen |
| Voice guidance | Voice prompts with adaptive volume; workflow adapts to the user’s pace |
| Languages | Up to six languages, selectable using the language button |
| Electrode shelf life | 60 months |
| Electrode cable / active area | 1.40 ±0.20 m cable; active electrode area 86 ±10 cm² |
| Maximum shocks per electrode set | 50 shocks |
| Electrode options | SavePads PLUS C (OBS-DE/P 303A1206), without CPR feedback sensor; SavePads PLUS CS (OBS-DE/P 303A1207), with a cable-connected CPR feedback sensor |
| Internal memory | 8 GB |
| Recording capacity | 160 hours of ECG waveform and CPR data; 10,000 events; at least 3,650 self-test reports |
| Data export | USB stick |
| Battery 3C — non-rechargeable | LiMnO₂; 12 V; nominal capacity 2.8 Ah; shelf life 60 months; 130 shocks at 200 J |
| Battery 3G — rechargeable option | Li-ion; 14.4 V; nominal capacity 2.95 Ah; shelf life 144 months; 230 shocks at 200 J |
| Dimensions — model 670A | 151 × 151 × 73 mm (height × width × depth; ±2 mm) |
| Weight — model 670A | Approximately 1.0 kg (±0.2 kg), including battery |
| Ingress protection | IP66 with the battery inserted and secured and the electrodes connected |
| Drop test | 1.6 m on all sides, corners and edges |
| Operating environment | 0°C to 55°C; 0–95% relative humidity, non-condensing; 540–1,060 hPa |
| Low-temperature operation | The manufacturer specifies at least 20 minutes of operation after transition from ambient temperature to −20°C |
| Storage environment | −5°C to 55°C; 0–95% relative humidity, non-condensing; 540–1,062 hPa |
| Transport environment — up to 10 days | −30°C to 70°C; 0–95% relative humidity, non-condensing; 510–1,062 hPa |
| Resuscitation guidelines | ERC 2021 |
| Manufacturer’s standard scope of delivery | HeartSave myPAD, one battery, one pair of electrodes, resuscitation kit and instructions for use |
The myPAD family data sheet also lists CPR feedback with SavePads PLUS CS (100–120 compressions/min, frequency accuracy ±3/min; depth 50–60 mm, accuracy ±5 mm or ±10%, whichever is larger), and optional LTE/Wi-Fi remote device management. These specifications depend on the selected accessories and configuration and do not confirm their inclusion with this product.
Source: PRIMEDIC HeartSave myPAD Technical data, document 24766, revision B, dated 18 June 2025. Model-specific dimensions, weight and LED guidance above apply to the 670A. Battery and electrode life depend on use and conditions. Refer to the manufacturer’s instructions for full operating details.